2600A系列数字源表实现精密脉冲定时和直流偏置开启序列

0
阅读(2649)

直流偏置开启序列的精密定时

Precision Timing of DC Bias Turn-On Sequence

 

在很多IC[1]ASIC[2]中,多个电源的开启序列至关重要。同样重要的是,不同电源开启之间的定时。在2600型系列数字源表中,不同电源开启之间延迟的精度受到连续命令行命令执行的限制。要将SMUA[3]SMUB[4]开启之间的延迟设置为200µs,可能需要以下命令序列:

In many ICs and ASICs,the turn-on sequence of multiple powersupplies is critical. Equally important is the timing betweenthe turn-on of the various supplies. In Series 2600 (non-A)instruments,the precision of the delay between the turn-on ofvarious power supplies was limited by the command executionof sequential command lines. The following command sequence might be used to program a 200µs delay between the turn-on ofSMUA and SMUB:

 

reset()smua.source.levelv = 0

smub.source.levelv = 0

smua.source.output = smua.OUTPUT_ON

smub.source.output = smub.OUTPUT_ON

smua.source.levelv = 5

delay(200e-6)

smub.source.levelv = 3.5

 

图3给出在2602型数字源表[5]上执行前述序列的结果。实际延迟接近1.5ms,比预设值多7倍。此外,对于连接至TSP-Link网络的每个节点,命令执行时间也都有所增加。

Figure 3shows the results of executing the precedingsequence on a Model 2602. The actual delay is nearly 1.5ms,more than 7 times the programmed value. Furthermore,thecommand execution time is multiplied for each node connectedto the TSP-Link network.

 

图3  2602型数字源表的直流偏置序列。期望延迟是200µs。实际延迟是1.45ms。

 

 

 了解更多信息
     要想了解有关2600A系列数字源表或者吉时利其他系列数字源表的更多信息,或者咨询全套直接可用的多通道测试系统的价格,请点击https://admin.acrobat.com/_a16893448/p81871518/,或者联系吉时利公司:全国免费电话800-810-1334手机用户请拨打440-650-1334。

登录吉时利官方微博(http://weibo.com/keithley)与专家进行互动。

 

吉时利仪器公司(中国)市场部
(T) 8008101334,  8610 82255010,  82255011

 

 
 


[1]IC http://www.keithley.com.cn/data?asset=50688

[2]ASIC http://www.keithley.com.cn/events/dianziyuanqijianceshi

[3]SMUA http://www.keithley.com.cn/promo/wb/294/download_file/Bipolar_Tran_App.pdf

[4]SMUB http://www.keithley.com.cn/images/promoimages/Ban_2600_04-a.swf/view_swf