2600A系列数字源表实现精密脉冲定时和直流偏置开启序列2
0赞2600A系列型数字源表[1]触发模型,允许用户在微秒数量级的精度对源电平和延迟时间进行编程。下面的命令序列将2602A[2]型数字源表SMUA[3]与SMUB[4]开启时间延迟设置为200µs 。
The Series 2600A trigger model allows the user topreprogram source levels and delay times with accuracies onthe order of microseconds. The following command sequenceprograms a 200µs delay between turn-on of SMUA and SMUB ona Model 2602A.
reset()
smua.trigger.source.listv({5})
smua.trigger.source.action = smua.ENABLE
smub.trigger.source.listv({3.5})
smub.trigger.source.action = smub.ENABLE
--Program timer with a delay of 200us
trigger.timer[1].delay = 200e-6
trigger.timer[1].stimulus = smua.trigger.ARMED_EVENT_ID
trigger.timer[1].count = 1
trigger.timer[1].passthrough = false
smua.trigger.source.stimulus = 0
smub.trigger.source.stimulus = trigger.timer[1].EVENT_ID
smua.trigger.endsweep.action = smua.SOURCE_HOLD
smub.trigger.endsweep.action = smub.SOURCE_HOLD
smua.source.output = smua.OUTPUT_ON
smub.source.output = smub.OUTPUT_ON
smub.trigger.initiate()
smua.trigger.initiate()
图4给出前述命令序列的结果。注意,通道之间的延迟大约是196.8µs(在50% SMUA振幅到50% SMUB振幅间测量)。
The results of the preceding command sequence are capturedin Figure 4. Note that the delay between the channels isapproximately 196.8µs (measured at 50% amplitude of SMUA to50% amplitude of SMUB).

图4 2602A型数字源表的直流偏置序列。期望延迟是200µs。实际延迟是196.8µs。
With Series 2600A instruments,a minimum delay time of10µs seconds is possible.
利用2600A系列数字源表,可以实现10µs的最小时间延迟。
Figure 4. DC bias sequencing with Model 2602. Desired delay was 200µs. Actual delay was 196.8µs.
了解更多信息
要想了解有关2600A系列数字源表或者吉时利其他系列数字源表的更多信息,或者咨询全套直接可用的多通道测试系统的价格,请点击https://admin.acrobat.com/_a16893448/p81871518/,或者联系吉时利公司:全国免费电话800-810-1334手机用户请拨打440-650-1334。
登录吉时利官方微博(http://weibo.com/keithley)与专家进行互动。
|
吉时利仪器公司(中国)市场部 |